PAM-01A1 series detectors are charged particles detector based on planar CZT crystal in a super small size. They have a high energy resolution in vacuum environment.
PAM-01A1 integrated custxomized CZT crystal and low noise charge sensitive preamplifier circuit. It can convert α-ray into exponential decay signals. With five-pin cable, high and low power supply, main amplifier,those signals will be direct into multi-channel pulse analyzer and then the energy spectrum will be formed.
1. CZT α-ray Probe Specification
Power | 0.3W |
Powered by | ±12W |
Output impedance | 50Ω |
Operating temperature | -20℃~40℃ (standard) / -20℃~70℃ (customized) |
High-voltage input | DC + 80V ~ 100V |
Effective window area | Φ25mm |
Signal output | exponential decay signal Tr<150ns |
Interface | Five-pin plug |
Energy resolution | <1%@5.48MeV @ room temperature |
8 hours peak drift | <0.2% |
2. CdZnTe α-ray Probe Spectrum
3. Features of CZT α-ray Detector
High energy resolution
Room temperature working;
Quick response
Working under visible light
4. Applications of CZT Probe with α-ray
Nuclear science & technology
Aerosol measurement
Hydrogen measurement
For more information, please contact us email at victorchan@powerwaywafer.com and powerwaymaterial@gmail.com.