α/β/γ-ray Differentiation Detector

α/β/γ-ray Differentiation Detector

PAM-02A2 series detectors are super small sized detectors based on planar CZT. They can detect α-ray, γ-ray and β-ray simultaneously.

PAM-02A2 integrates CZT crystal, low noise charge preamplifier circuit, SK shaping circuit and high-voltage circuit. With ordered cable, it can output Quasi-gaussian signal or digital signal for back-end data processing unit.

Referred to the graph, α-ray andβ-ray are mainly act on CZT1. But the energy of α-ray is far higher than β-ray. So we can tell the two rays according to their amplitude. γ-ray energy will deposit both in CZT1 and CZT2, and there energy deposit are seemingly even. So through comparing these in the two detectors, they can tell γ-ray. At the same time, it will filter out the interference of signals from β-ray.

Specification

Power 0.1W
Powered by ±5W
Output impedance 50Ω
Operating temperature -20℃-+40℃
High voltage Built-in high-voltage
Crystal type Planar CZT detector 20*20*0.5mm3
Signal output 0-5VQuasi-Gaussian pulse or TTL
Intrinsic efficiency α:46%@241Am
β:29%@90Sr
γ:2%@137Cs

Spectrum

 

 

 

 

 

 

 

Features
Highly sensitive
Room temperature working

Applications
Environment system
Security system

 